Product Details
FAB
- Rapid - Results can be obtained within 90 minutes.
- Reference Standard Traceability - The reference standard is traceable to the National Standard USP (Cat # 1076192).
- Rigorous Method Validation - Validation reports compliant with ICH guidelines can be provided upon request.
- High Sensitivity - Sensitivity for Bovine Serum Albumin (BSA) is < 0.5 ng/mL.
- Strong Specificity - No cross-reactivity with non-target proteins.
- Excellent Buffer Compatibility - Compatible with a wide range of buffer systems.
Product Specifications
Assay TypeSandwich-ELISAAnalyteBSA (Bovine serum albumin)Format96T (8×12 strips)Regulatory StatusRUOSensitivity<0.5 ng/mLStandard Curve Range0.5 ng/mL-40.5 ng/mLAssay Time1.5 hrSuitable Sample TypeFor the quantitative determination of BSA (Bovine serum albumin)Sample volume50 μLTraceabilityUSP (Cat # 1076192)Product Overview
resDetect™ BSA ELISA Kit (USP Calibrated) is based on the ELISA sandwich method and is used to detect and quantitatively determine Bovine Serum Albumin (BSA) residues in biological products. This kit is designed to support optimal purification process development and routine quality control of in-process streams and final products in various biotechnological processes, such as cell and tissue culture. The standard BSA has been calibrated against the USP standard (cat # 1076192) to ensure accurate assay results. It can also be used as a universal detection tool for the quantitative determination of BSA.
Storage
1. Unopened kit should be stored at 2℃-8℃ upon receiving.
2. Find the expiration date on the outside packaging and do not use reagents past their expiration date.
3. The opened kit should be stored per components table. The shelf life is 30 days from the date of opening.
Materials Provided
IDComponentsSizeRES98-C01Pre-Coated Anti-BSA Antibody Microplate1 plate (8×12 strips)RES98-C02ABSA Standard (200ng/mL)500 μLRES98-C02BBSA Standard 5 (40.5ng/mL)500 μLRES98-C02CBSA Standard4 (13.5ng/mL)500 μLRES98-C02DBSA Standard 3 (4.5ng/mL)500 μLRES98-C02EBSA Standard 2 (1.5ng/mL)500 μLRES98-C02FBSA Standard 1 (0.5ng/mL)500 μLRES98-C02GBSA Standard 0 (0ng/mL)500 μLRES98-C03HRP-Anti-BSA Antibody100 μLRES98-C041×Sample Dilution Buffer50 mL×2RES98-C0520×Washing Buffer25 mLRES98-C06Substrate Solution12 mLRES98-C07Stop Solution8 mLACRO Quality Management System
Customers Also Viewed
Performance Data
Typical Data
Please refer to DS document for the assay protocol.

Detection of BSA (Bovine Serum Albumin) by ELISA Assay.
Immobilized Anti-BSA Antibody can bind BSA (Bovine Serum Albumin) Detection was performed using HRP-Anti- BSA Antibody with sensitivity of 0.5 ng/mL (QC tested). For each experiment, a standard curve needs to be set for each micro-plate, and the specific OD value may vary depending on different laboratories, testers, or equipments. The following example data is for reference only.
Dilution Linearity
To assess the linearity of the assay, samples spiked with high concentrations of BSA were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.

Intra-Assay Statistics
Three samples of known concentration were tested ten times on one plate to assess intra-assay precision , Intra-Assay Precision CV≤15%.

Inter-Assay Statistics
Three samples of known concentration were tested in three separate assays to assess inter-assay precision, Inter-Assay Precision CV≤15%.

Accuracy
Five samples of known concentration were tested ten times to assess Accuracy , Accuracy recovery rate 80-120%.

Interference effect
Verify potential matrix effects by adding different levels of specific buffers to the sample buffer. For specific buffers, it is recommended that you verify recovery to determine the optimal dilution ratio.

Specificity
High, medium and low concentrations of BSA were added to Human serum (v/v 1%), Mouse serum (v/v 2.5%), Human Serum Albumin (1.25mg/mL) and Rabbit serum (v/v 10%), the recovery rate of BSA was used as the specific validation index.

High, medium and low concentrations of BSA were added to MDCK Cell, HEK293 Cell, CHO Cell , T Cell and VERO Cell, the recovery rate of BSA was used as the specific validation index.

Customer Reviews Writing Reviews